FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS

MDR report key: 2921344 · Received January 17, 2013

Report

Report Number
2955842-2013-00231
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 18, 2012
Report Date
December 19, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI HYSTERECTOMY PROCEDURE, WHILE USING THE PK DISSECTING FORCEPS INSTRUMENT, THE CUSTOMER NOTED A BROKEN CABLE AT THE TIP. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.