FDA Adverse Event Malfunction Summary report: N

PHYSIO CONTROL

MDR report key: 292114 · Received August 8, 2000

Report

Report Number
MW4002817
Event Type
Malfunction
Date Received
August 8, 2000
Date of Event
July 20, 2000
Report Date
July 31, 2000
Manufacturer
PHYSIO CONTROL
Product Code
MLN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TIP OF FAST PATCH CONNECTOR POST BROKE OFF INSIDE CONNECTOR OF PHYSIO CONTROL LIFEPAK 10. THIS PREVENTED USER FROM GETTING SOLID CONNECTION. ALTHOUGH SOME ARCING OCCURRED WITHIN CONNECTOR, PT WAS REVIVED WITHOUT FURTHER INTERVENTION. CONCERN IS THAT BAD CONNECTION COULD RESULT IN ADVERSE EVENT UNDER DIFFERENT CIRCUMSTANCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHYSIO CONTROL DEFIBRILLATOR FAST-PATCH MLN PHYSIO CONTROL QUANTUM EDGE FAST-PATCH NI

Patients

Seq Age Sex Outcome Treatment
1 49 YR