FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2921019 · Received January 17, 2013

Report

Report Number
1416980-2013-01307
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
October 29, 2012
Report Date
December 28, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR; TIDAL TOTAL ULTRA-FILTRATION (UF) REMOVAL WAS SET TOO LOW. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012 AT 22:20:58. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 2196ML, INDICATING THE HOME PATIENT (HP) DRAINED 1896ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 3000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26552 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 30 YR