FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 2920921 · Received January 17, 2013

Report

Report Number
2124215-2013-01086
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 7, 2012
Report Date
May 9, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT PACEMAKER WAS DISPLAYING VARYING PROGRAMMER LONGEVITY ESTIMATES. AUTOMATIC CAPTURE WAS PROGRAMMED TO OFF IN AN EFFORT TO REDUCE VARIABILITY. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE DEVICE REMAINS IN SERVICE.

Description of Event or Problem · 1

SUBSEQUENT INFORMATION INDICATES THAT THE DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THE DEVICE HAS BEEN RETURNED FOR ANALYSIS. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26140 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1291

Patients

Seq Age Sex Outcome Treatment
1 93 YR 5076| 1291