FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2920885
·
Received December 26, 2012
Report
- Report Number
- 1824206-2012-08504
- Event Type
- Malfunction
- Date Received
- December 26, 2012
- Date of Event
- December 6, 2012
- Report Date
- December 6, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH REPLACED THE EMERGENCY TRENDELENBURG VALVE AND THE HEAD HI-LO DOWN VALVE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE TECH ALLEGED THE HEAD HI-LO WOULD DRIFT SLOWLY. NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED TOTAL CARE | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |