FDA Adverse Event Malfunction Summary report: N

LIGACLIP*MCA LARGE APPLIER

MDR report key: 2920793 · Received January 17, 2013

Report

Report Number
3005075853-2013-00268
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
October 24, 2012
Report Date
January 10, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDO
PMA / PMN Number
K820837
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. ADDITIONAL INFORMATION: MULTIPLE REQUEST FOR RETURN OF DEVICE HAVE BEEN MADE BUT NO DEVICE HAS BEEN RETURNED.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WERE DROPPING CLIPS. THE CLIPS DROPPED FROM THE DEVICE BEFORE CLOSURE OF THE HANDLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25596 LIGACLIP*MCA LARGE APPLIER CLIP, IMPLANTABLE GDO ETHICON ENDO-SURGERY, LLC. NA J4C19D

Patients

Seq Age Sex Outcome Treatment
1