ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-00822
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Report Date
- December 22, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/05/2014 WITH THE FOLLOWING FINDINGS: NO DEFECT WAS FOUND. NO BATTERY CAP RETURNED. ALL TESTING PERFORMED WITH A TEST BATTERY CAP. PUMP POWERS WITH A TEST BATTERY CAP TO A DIM DISCOLORED DISPLAY. BATTERY CAP UNABLE TO FULLY TIGHTEN. NO POWER LOSS WAS OBSERVED. PUMP WAS EXERCISED WITH A TEST CAP FOR 24 HOURS WITH NO REBOOT, LOSS OF POWER OR CALL SERVICE ALARMS DUPLICATED. CURRENT DRAWS WITHIN SPECIFICATIONS. REMOVED PUMP CASE. NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS INSIDE PUMP 6. BLACK BOX DATES FROM (B)(6) 2014. COMPLAINT WAS LOGGED ON 12/22/2012. NO UNEXPLAINABLE POWER ON RESET EVENTS OBSERVED IN CURRENT BLACK BOX.
ON (B)(4) 2012 THE REPORTER CONTACTED ANIMAS AND ALLEGED HIS PUMP IS RANDOMLY SHUTTING DOWN ON HIM. NO OTHER INFORMATION IS PROVIDED. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGATION THAT THE PUMP IS INTERMITTENTLY LOSING POWER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25954 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |