FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC

MDR report key: 2920706 · Received December 10, 2012

Report

Report Number
1824206-2012-08027
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 19, 2012
Report Date
November 19, 2012
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE CABLE SHIELD WAS CATCHING THE LOCKING MECHANISM. HE BENT THE SHIELD BACK INTO POSITION TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATED THAT THE RIGHT SIDERAIL WOULD NOT LATCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BARIATRIC AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1840

Patients

Seq Age Sex Outcome Treatment
1