FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2920634 · Received December 7, 2012

Report

Report Number
3008642652-2012-03197
Event Type
Malfunction
Date Received
December 7, 2012
Date of Event
November 20, 2012
Report Date
December 3, 2012
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CONSTANT CHECK BELT AND CHECK TE PAD ALARMS) HAS BEEN CONFIRMED. UPON EVALUATION, THERE WAS AN OPEN PULSE WIRE INSIDE THE CABLE CONNECTING ECG ELECTRODES A AND B. THE CAUSE OF THE CHECK BELT AND CHECK TE PAD ALARMS IS THE DAMAGED PULSE WIRE. THE ROOT CAUSE OF THE DAMAGED WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK BELT AND CHECK TE PAD ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 31 YR