FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2920634
·
Received December 7, 2012
Report
- Report Number
- 3008642652-2012-03197
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Date of Event
- November 20, 2012
- Report Date
- December 3, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CONSTANT CHECK BELT AND CHECK TE PAD ALARMS) HAS BEEN CONFIRMED. UPON EVALUATION, THERE WAS AN OPEN PULSE WIRE INSIDE THE CABLE CONNECTING ECG ELECTRODES A AND B. THE CAUSE OF THE CHECK BELT AND CHECK TE PAD ALARMS IS THE DAMAGED PULSE WIRE. THE ROOT CAUSE OF THE DAMAGED WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK BELT AND CHECK TE PAD ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |