FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2920619 · Received January 17, 2013

Report

Report Number
1416980-2013-01281
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
October 18, 2012
Report Date
October 24, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RS
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS CONDITION IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF ANOTHER CONDITION. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO THE NEXT FILL WHEN A SLOW / NO FLOW OCCURRED, ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012. DURING NIGHT DRAIN CYCLE 14, THE PATIENT'S ULTRAFILTRATION READING WAS 884ML. THIS INDICATES THAT THE HOME PATIENT (HP) DRAINED 704ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 900ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26340 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO