LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-00265
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Date of Event
- January 8, 2013
- Report Date
- January 10, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON?---FIRST. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? ---BLOOD VESSEL. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? ---NO INFORMATION. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? ---NO. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? YES. WERE ANY UNEXPECTED NOISES HEARD? UNK. IF SO, WHEN? DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? ---NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? YES, OUTSIDE. WHAT WERE THE INDICATIONS FOR SURGERY? GALL BLADDER DISEASE. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? ---NO INFORMATION. DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? IF SO, WHOM? ---NO. INFORMATION THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION WITH ONE UNFORMED AND TWO GAP CLIPS INSIDE A PLASTIC BAG. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIP AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THERE WAS AN UNEXPECTED RESISTANCE AT THE FIRING. WHEN THE DEVICE WAS FIRED OUTSIDE THE PATIENT, THE CLIP WAS MALFORMED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25876 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |