FDA Adverse Event Malfunction Summary report: N

VOYAGER NC CORONARY DILATATION CATHETER

MDR report key: 2920572 · Received January 17, 2013

Report

Report Number
2024168-2013-00325
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 26, 2012
Report Date
December 26, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110617
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED BALLOON RUPTURE WAS CONFIRMED. BASED ON VISUAL AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A 90% STENOSED LESION IN THE MID LEFT ANTERIOR DESCENDING ARTERY WITH MODERATE TORTUOSITY AND MILD CALCIFICATION. THE 2.75 X 12 MM VOYAGER NC BALLOON CATHETER WAS ADVANCED WITHOUT ISSUE FOR POST-DILATATION; HOWEVER, DURING AN INFLATION TO 14 ATMOSPHERES (ATM), A BALLOON RUPTURE OCCURRED. IT IS UNKNOWN HOW MANY INFLATIONS WERE MADE WITH THE VOYAGER NC. THE BALLOON CATHETER WAS REMOVED WITHOUT ISSUE AND A NEW VOYAGER NC WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25638 VOYAGER NC CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 1042061

Patients

Seq Age Sex Outcome Treatment
1