FDA Adverse Event Malfunction Summary report: N

MUSTANG¿

MDR report key: 2920322 · Received January 17, 2013

Report

Report Number
2134265-2013-00066
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 18, 2012
Report Date
December 21, 2012
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
LIT
PMA / PMN Number
K103751
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE DEVICE WAS RECEIVED WITH A 0.035INCH GUIDEWIRE TRAPPED INSIDE THE DEVICE. THE GUIDEWIRE WAS SEVERELY STRETCHED AT THE SECTION EXITING THE WIRE LUMEN AT THE HUB OF THE DEVICE. AN ATTEMPT TO REMOVE THE WIRE FROM THE DEVICE WAS MET WITH RESISTANCE. AS A RESULT THE SHAFT WAS DISSECTED AT VARIOUS LOCATIONS AND EVENTUALLY THE GUIDEWIRE WAS REMOVED. NO ANOMALIES WERE NOTED. THE BALLOON IS PARTIALLY INFLATED WITH CONTRAST MEDIA. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS UNABLE TO BE DETERMINED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE THE BALLOON STUCK ON THE WIRE. THE DETAILS OF THE LESION ARE UNKNOWN. THE 7.0 X 80, 75CM MUSTANG BALLOON CATHETER BECAME STUCK ON A NON-BSC GUIDEWIRE. DETAILS OF HOW THE PROCEDURE WAS COMPLETED IS UNKNOWN. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS OK.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE THE BALLOON STUCK ON THE WIRE. THE DETAILS OF THE LESION ARE UNKNOWN. THE 7.0 X 80, 75CM MUSTANG BALLOON CATHETER BECAME STUCK ON A NON-BSC GUIDEWIRE. DETAILS OF HOW THE PROCEDURE WAS COMPLETED IS UNKNOWN. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS OK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26033 MUSTANG¿ CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - GALWAY H74939171070870 15683997

Patients

Seq Age Sex Outcome Treatment
1