FDA Adverse Event Malfunction Summary report: N

COYOTE¿ ES

MDR report key: 2920306 · Received January 17, 2013

Report

Report Number
2134265-2012-08561
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 21, 2012
Report Date
December 21, 2012
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 100% STENOSED LESION WAS LOCATED IN A MODERATELY TORTUOUS AND CALCIFIED ANTERIOR TIBIAL ARTERY. USING A NON-BSC SHEATH AND GUIDEWIRE, THE 4MM X 20MM X 144CM COYOTE ES BALLOON CATHETER WAS ADVANCED AND BEING USED TO DILATE THE LESION. THE BALLOON WAS INFLATED TO 6 ATMS THEN TO 12 ATMS AND RUPTURED AT 12 ATMS UPON THIRD INFLATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26472 COYOTE¿ ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939135402010 14565441

Patients

Seq Age Sex Outcome Treatment
1