FDA Adverse Event Malfunction Summary report: N

UNIVERSAL DRILL

MDR report key: 2920221 · Received December 17, 2012

Report

Report Number
1811755-2012-04272
Event Type
Malfunction
Date Received
December 17, 2012
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS NOTED DURING FAILURE ANALYSIS THAT THE CONTACT PINS ARE BURNT AND BENT. CORROSION DAMAGE AND VARNISH BUILD-UP WAS NOTED WITHIN THE INTERNAL COMPONENTS OF THE DEVICE.

Description of Event or Problem · 1

THE UNIVERSAL DRILL WAS SENT FOR EVALUATION DUE TO RUNNING ON ITS OWN WITHOUT USER ACTIVATION DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT ANY PROCEDURE DELAY. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNIVERSAL DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK