FDA Adverse Event
Malfunction
Summary report: N
UNIVERSAL DRILL
MDR report key: 2920221
·
Received December 17, 2012
Report
- Report Number
- 1811755-2012-04272
- Event Type
- Malfunction
- Date Received
- December 17, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 16, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS NOTED DURING FAILURE ANALYSIS THAT THE CONTACT PINS ARE BURNT AND BENT. CORROSION DAMAGE AND VARNISH BUILD-UP WAS NOTED WITHIN THE INTERNAL COMPONENTS OF THE DEVICE.
Description of Event or Problem · 1
THE UNIVERSAL DRILL WAS SENT FOR EVALUATION DUE TO RUNNING ON ITS OWN WITHOUT USER ACTIVATION DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT ANY PROCEDURE DELAY. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL DRILL | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |