CORE IMPACTION DRILL
MDR report key: 2920218
·
Received December 17, 2012
Report
- Report Number
- 1811755-2012-04281
- Event Type
- Malfunction
- Date Received
- December 17, 2012
- Date of Event
- November 20, 2012
- Report Date
- November 26, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVALUATION, THE BEARINGS AND THE ROTOR WERE FOUND TO BE CORRODED, WHICH IS A PROBABLE CAUSE OF OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A WISDOM TOOTH REMOVAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO CLINICALLY SIGNIFICANT DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.