ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-00798
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Report Date
- January 8, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, FR
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 02/13/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE KEYPAD WAS PEELING AT THE UP ARROW KEYPAD BUTTON. A DAMAGED KEYPAD WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD SHOULD BE CLEARLY VISIBLE AND WARNS THE PATIENT TO DISCONTINUE USING THE PUMP. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED. EVALUATION REVEALED THAT THE OK AND DOWN ARROW KEYPAD BUTTONS ARE INTERMITTENTLY UNRESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. THE OK BUTTON KEY CONTACT IS INVERTED.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE OK KEYPAD BUTTON HAD A DELAYED RESPONSE AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE FOR TWO WEEKS. THE OK BUTTON WAS NOW STICKING AND CAUSED SCROLLING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25569 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |