FDA Adverse Event Injury Summary report: N

STANDARD IM NAIL

MDR report key: 2920091 · Received January 8, 2013

Report

Report Number
3034525-2013-00002
Event Type
Injury
Date Received
January 8, 2013
Date of Event
January 2, 2013
Report Date
January 8, 2013
Manufacturer
SIGN FRACTURE CARE INTERNATIONAL
Product Code
HSB
PMA / PMN Number
K022632
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

DATABASE REPORT OF A BROKEN IM NAIL, THE SURGEON INDICATED AN EXCHANGE NAIL PROCEDURE TO CHANGE THE BROKEN 8MM NAIL FOR A 10MM NAIL. REVIEW OF THE PATIENT X-RAYS SHOWS A NAIL BROKEN AT ONE OF THE DISTAL SCREW SLOTS. THE X-RAYS ALSO SHOWED A NON-UNION WHICH APPEARS TO HAVE SHIFTED CAUSING STRESS AT THE DISTAL END OF THE NAIL. CAUSE: NON-UNION SHIFTED CAUSING STRESS TO BE PLACED ON THE DISTAL END OF THE IM NAIL AND CAUSING THE NAIL TO BREAK AT A WEAK POINT (SCREW SLOT). NO INDICATION OF PRODUCT DEFECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10461 STANDARD IM NAIL STANDARD IM NAIL HSB SIGN FRACTURE CARE INTERNATIONAL STANDARD IM NAIL UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 36 YR Hospitalization