STANDARD IM NAIL
MDR report key: 2920091
·
Received January 8, 2013
Report
- Report Number
- 3034525-2013-00002
- Event Type
- Injury
- Date Received
- January 8, 2013
- Date of Event
- January 2, 2013
- Report Date
- January 8, 2013
- Manufacturer
- SIGN FRACTURE CARE INTERNATIONAL
- Product Code
- HSB
- PMA / PMN Number
- K022632
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
DATABASE REPORT OF A BROKEN IM NAIL, THE SURGEON INDICATED AN EXCHANGE NAIL PROCEDURE TO CHANGE THE BROKEN 8MM NAIL FOR A 10MM NAIL. REVIEW OF THE PATIENT X-RAYS SHOWS A NAIL BROKEN AT ONE OF THE DISTAL SCREW SLOTS. THE X-RAYS ALSO SHOWED A NON-UNION WHICH APPEARS TO HAVE SHIFTED CAUSING STRESS AT THE DISTAL END OF THE NAIL. CAUSE: NON-UNION SHIFTED CAUSING STRESS TO BE PLACED ON THE DISTAL END OF THE IM NAIL AND CAUSING THE NAIL TO BREAK AT A WEAK POINT (SCREW SLOT). NO INDICATION OF PRODUCT DEFECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10461 | STANDARD IM NAIL | STANDARD IM NAIL | HSB | SIGN FRACTURE CARE INTERNATIONAL | STANDARD IM NAIL | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Hospitalization |