FDA Adverse Event Injury Summary report: N

MPS DELIVERY SET W AA AND ADD CASSETTES

MDR report key: 2920089 · Received December 14, 2012

Report

Report Number
1649914-2012-00036
Event Type
Injury
Date Received
December 14, 2012
Date of Event
May 1, 2011
Report Date
May 12, 2011
Manufacturer
QUEST MEDICAL, INC.
Product Code
DTR
PMA / PMN Number
K953838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION - DEVICE EVAL FOUND THE ROOT CAUSE TO BE INSUFFICIENT SOLVENT-BONDING WHERE THE BLOOD SOURCE TUBE-LINE CONNECTS TO THE BUSHING OF THE PUMP CASSETTE. A CORRECTIVE ACTION HAS BEEN IDENTIFIED AND IMPLANTED TO ADDRESS THE ROOT CAUSE. QUEST MEDICAL, INC HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. QUEST MEDICAL, INC DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED BY THE HOSP PERFUSIONIST THAT DURING A PROCEDURE, THE DELIVERY SET DISPOSABLE EXPERIENCED A LEAK IN THE TUBING AT THE CONNECTION POINT. THE PERFUSIONIST REPORTED THE PT LOST APPROX 100 CCS OF BLOOD DUE TO THE ALLEGED MALFUNCTION. IT WAS REPORTED THAT THE TUBING WAS CINCHED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PT COMPLICATIONS REPORTED. THE DISPOSABLE DEVICE WAS RETURNED TO THE MFR FOR ANALYSIS. NO ADD'L INFO IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MPS DELIVERY SET W AA AND ADD CASSETTES CPBP HEAT EXCHANGER DTR QUEST MEDICAL, INC. 5001102 0393451M12

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention