FDA Adverse Event Malfunction Summary report: N

9F PRUITT OUTLYING F3 SHUNT WITH T-PORT

MDR report key: 2917876 · Received December 20, 2012

Report

Report Number
1220948-2012-00023
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
December 10, 2012
Report Date
December 10, 2012
Manufacturer
LEMAITRE VASCULAR
Product Code
MJN
PMA / PMN Number
K051067
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RETURNED FOR THE EVALUATION AND, WE WERE ABLE TO CONFIRM THE FAILURE. WE FOUND THAT THE BLUE STOPCOCK ON THE SHUNT WAS LEAKING AT THE BUSHING. THE MOST PROBABLE ROOT CAUSE OF LEAKING IS THE OPERATOR ERROR ¿ NOT ENOUGH GLUE WAS APPLIED DURING THE MANUFACTURING PROCESS. THE DEVICE HISTORY RECORD REVIEW DID NOT REVEAL ANY DISCREPANCIES RELATED TO THE COMPLAINT EVENT DURING EITHER THE MANUFACTURING OR THE PACKAGING PROCESSES. WE HAVE INITIATED CORRECTIVE AND PREVENTIVE ACTION (PLEASE REFERENCE LEMAITRE VASCULAR, INC. CAPA (B)(4)) TO ADDRESS THIS ISSUE IN MORE DETAILS. WE DID MAKE OUR MANUFACTURING ASSOCIATES AWARE OF THIS PROBLEM AND RE-TRAINED THEM. PLEASE NOTE THAT ALL OUR DEVICES ARE 100% LEAK (INFLATION) TESTED BEFORE THE SHIPPING. HOWEVER, WE ARE INVESTIGATING LEAK TEST METHODS. PLEASE NOTE THAT NO PATIENT INJURY HAPPENED.

Description of Event or Problem · 1

SHUNT FAILED WHILE SURGEON WAS SEWING IN PATCH AT THE END OF THE PROCEDURE, BUT WORKED UP UNTIL THAT POINT. A BLUE COMMON BALLOON KEPT SLIPPING OUT OF THE ARTERY. THE PHYSICIAN FOUND A LEAK RIGHT BETWEEN BLUE STOPCOCK WHERE THE SYRINGE WAS ATTACHED AND TUBING. SHUNT WORKED PROPERLY DURING PRE USE TEST AND THEN THE FIRST HALF OF THE PROCEDURE. PATIENT SAFETY WAS COMPROMISED DUE TO PROLONGED PROCEDURE: THEY HAD TO CLAMP FOR 10 MINUTES AND THEN ANOTHER 3 MINUTES. HOWEVER, NO PATIENT INJURY HAPPENED.