PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2013-00111
- Event Type
- Malfunction
- Date Received
- January 15, 2013
- Date of Event
- December 20, 2012
- Report Date
- December 20, 2012
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, PRIME AND DISPLACEMENT TEST. HOWEVER, THE INSULIN PUMP WAS STUCK IN MOTOR ERROR LOOP DURING BOLUS/BASAL DELIVERY. ERROR ALARM CONFIRMED IN THE HISTORY FILE. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING.
THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING NORMAL USE. THE CUSTOMER STATED THAT THE INSULIN PUMP HAS NOT BEEN EXPOSED TO HIGH MAGNETIC FIELDS. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 22860 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |