FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2915366
·
Received January 9, 2013
Report
- Report Number
- 8020893-2013-00041
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 17, 2012
- Report Date
- December 20, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD AN ERRATIC DISPLAY. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHICAL USER INTERFACE (GUI) PCB. THE VENTILATOR PASSED ALL TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12621 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |