FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2915308 · Received January 9, 2013

Report

Report Number
8020893-2013-00054
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. NO PT INVOLVEMENT. THE CSE INSPECTED THE DEVICE AND REPLACED THE BACKLIGHT INVERTER PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12455 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1