FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2915299 · Received December 19, 2012

Report

Report Number
3004123209-2012-00149
Event Type
Malfunction
Date Received
December 19, 2012
Date of Event
September 14, 2012
Report Date
December 13, 2012
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K052465
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2010, AND THAT IT HAD PERFORMED TO SPECIFICATION UP TO (B)(6) 2012. THE INFO OBTAINED FROM THE DEVICE SHOWED THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY RESULTING IN MANUAL POWER-UPS OF 10 MINUTES IN DURATION OCCURRING ON (B)(6) 2012, AND CONTINUED CONSISTENTLY UP TO (B)(6) 2012. INVESTIGATION DID NOT CONFIRM A FAULT WITH THE DEVICE OR ANY COMPONENT IN THE DEVICE. THE DEVICE SWITCHING ITSELF ON IS A KNOWN SYMPTOM OF A DAMAGED OR FAULTY MEMBRANE. ALTHOUGH IN THIS EVENT, THERE WAS NO FAULT MEASURED, IT IS PROBABLE THAT DAMAGE HAS BEEN CAUSED TO THE MEMBRANE, WHICH HAS AFFECTED THE DEVICE CAUSING IT TO SWITCH ITSELF ON AUTOMATICALLY. THIS CAN CAUSE PREMATURE DEPLETION OF THE PAD-PAK. THE RETURNED PAD-PAK (LOT 691), HAD A USE UNTIL DATE OF 07/2012, BUT WAS FOUND TO HAVE BEEN SIGNIFICANTLY DEPLETED. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE, BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD PAD

Patients

Seq Age Sex Outcome Treatment
1