FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2915230 · Received December 20, 2012

Report

Report Number
3004123209-2012-00259
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
October 31, 2012
Report Date
November 26, 2012
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K052465
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2010 AND THAT IT HAD PERFORMED TO SPEC UP TO (B)(6) 2012. THE INFO OBTAINED FROM THE DEVICE SHOWED THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY RESULTING IN MANUAL POWER-UPS OF 10 MINUTES IN DURATION OCCURRING ON (B)(6) 2012 AND CONTINUED CONSISTENTLY UP TO (B)(6) 2012. ON THE (B)(6) 2012, THE DEVICE FAILED A WEEKLY MANUAL POWER-UP BECAUSE OF A LOW BATTERY. THE DEVICE WOULD HAVE SWITCHED TO FAULT MODE AND THE CUSTOMER WOULD HAVE BEEN ALERTED BY THE STATUS INDICATOR FLASHING RED AND THE DEVICE EMITTING AN AUDIBLE WARNING MESSAGE. INVESTIGATION CONFIRMED THERE TO BE A FAULT WITH THE DEVICE MEMBRANE. THIS FAULT CAUSED THE DEVICE TO SWITCH ON AUTOMATICALLY, WHICH CAUSED THE EARLY DEPLETION OF THE PAD PAK (LOT 667). PAD PAK (LOT 667) HAD A USE UNTIL DATE OF 06/2013 BUT BECAME DEPLETED ON (B)(6) 2012. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD PAD

Patients

Seq Age Sex Outcome Treatment
1