HEARTSINE SAMARITAN 300 P AND PAD PAK
Report
- Report Number
- 3004123209-2012-00264
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- November 5, 2012
- Report Date
- November 28, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NOT APPLICABLE
Narratives
ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2009 AND THAT IT HAD PERFORMED TO SPECIFICATION UP TO (B)(6) 2012. THE INFO OBTAINED FROM THE DEVICE SHOWED THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY RESULTING IN MANUAL POWER-UPS OF 10 MINUTES IN DURATION OCCURRING ON (B)(6) 2012. ON (B)(6) 2012, THE DEVICE FAILED A WEEKLY SELF-TEST BECAUSE OF A LOW BATTERY AND SWITCHED TO FAULT MODE, WHICH THE CUSTOMER WOULD HAVE BEEN ALERTED TO WITH THE STATUS INDICATOR FLASHING RED AND THE DEVICE EMITTING AN AUDIBLE WARNING MESSAGE. THE LOW BATTERY WARNINGS BECAME CRITICAL AND THE PAD-PAK BECAME FULLY DEPLETED ON (B)(6) 2012. INVESTIGATION CONFIRMED THERE TO BE A FAULT WITH THE DEVICE MEMBRANE. THIS FAULT CAUSED THE DEVICE TO SWITCH ON AUTOMATICALLY, WHICH CAUSED THE EARLY DEPLETION OF THE PAD PAK (LOT A1220). PAD PAK (LOT A1220) HAD A USE UNTIL DATE OF 06/2016 BUT BECAME DEPLETED ON (B)(6) 2012. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSINE SAMARITAN 300 P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |