HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2012-00664
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- May 31, 2012
- Report Date
- November 3, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2008 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. ON (B)(6) 2011 THE DEVICE FAILED A SELF-TEST DUE TO A LOW BATTERY. THE USER WAS ALERTED TO THE PROBLEM BY THE RED LED STATUS INDICATOR BEING ILLUMINATED AND AN AUDIBLE BEEP. A NEW PAD-PAK WAS INSERTED AND THE DEVICE PERFORMED TO SPECIFICATION UNTIL (B)(6) 2012. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE STATUS INDICATOR WAS FLASHING AND THE DEVICE WAS EMITTING A LOW BATTERY WARNING. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED IF REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |