HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2012-00773
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- August 15, 2010
- Report Date
- December 11, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2009 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2010. AFTER THIS DATE THERE ARE MULTIPLE EVENTS ON THE SAME DAY WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO AN EXCESSIVE SOLDER WHICH WAS CAUSING AN INTERMITTENT CONNECTION BETWEEN TWO COMPONENTS IN THE SWITCHING CIRCUITRY OF THE DEVICE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.