FDA Adverse Event Injury Summary report: N

CATALYS PRECISION LASER SYSTEM

MDR report key: 2915067 · Received January 10, 2013

Report

Report Number
3005675890-2013-00002
Event Type
Injury
Date Received
January 10, 2013
Date of Event
December 11, 2012
Report Date
January 10, 2013
Manufacturer
OPTIMEDICA CORPORATION
Product Code
OOE
PMA / PMN Number
K121091
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION OF THE INCIDENT INCLUDED THE ANALYSIS OF THE SYSTEM DATABASE, SYSTEM OPTICAL COHERENCE TOMOGRAPHY (OCT) RECORDING, SYSTEM VIDEO DISPLAY RECORDING AND OPERATING ROOM SURGICAL VIDEO. FORM THE ANALYSIS OF THE SYSTEM DISPLAY RECORDING AND THE SYSTEM DATABASE THERE WERE NO ANOMALIES AND ALL SETTINGS AND PARAMETERS OF THE SYSTEM WERE FOUND TO BE WITHIN ACCEPTABLE LIMITS. FORM THE ANALYSIS OF THE SURGICAL VIDEO IT WAS NOTED THE SURGEON DID NOT USE CONTINUOUS CURVILINEAR CAPSULORRHEXIS (CCC) TECHNIQUE TO EXTRACT THE CAPSULE IN A CIRCUMFERENTIAL FASHION AND THAT THIS MAY HAVE CAUSED AND/OR CONTRIBUTED TO THE ANTERIOR CAPSULE TEAR. THE SPECIFIC CAUSE(S) OF THE ANTERIOR LENS CAPSULE TEAR ARE UNKNOWN. THE CATALYST SYSTEM OPERATOR MANUAL CONTAINS A WARNING WHICH STATES: "STANDARD CONTINUOUS CURVILINEAR CAPSULORRHEXIS (CCC) SURGICAL TECHNIQUE MUST BE USED FOR SURGICAL REMOVAL OF THE CAPSULOTOMY DISC. THE CAPSULOTOMY MAY HAVE RESIDUAL UNCUT AREAS THAT SHOULD BE COMPLETED BY ADVANCING THE CAPSULE THROUGH THE INCOMPLETELY CUT AREA IN A CIRCUMFERENTIAL FASHION, RATHER THAN PULLING IT RADIALLY. THE USE OF IMPROPER CAPSULOTOMY DISC REMOVAL TECHNIQUE MAY POTENTIALLY CAUSE OR CONTRIBUTE TO ANTERIOR CAPSULE TEAR AND/OR A NONCIRCULAR, IRREGULARLY SHAPED CAPSULOTOMY."

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT WHO UNDERWENT ANTERIOR CAPSULOTOMY, LENS FRAGMENTATION, AND CORNEAL INCISIONS WITH THE CATALYS SYSTEM (SYSTEM) SUBSEQUENTLY EXPERIENCED AN ANTERIOR LENS CAPSULE TEAR DURING THE SURGICAL PROCEDURE TO REMOVE THE CATARACT. IT WAS NOTED THAT THE PHYSICIAN DID NOT USE THE OPTIMREDICA RECOMMENDED CONTINOUS CURVILINEAR CAPSULORRHEXIS (CCC) CAPSULOTOMY DISC REMOVAL TECHNIQUE TO EXTRACT THE CAPSULOTOMY DIS. THE SURGEON NOTICED THE TEAR AFTER THE COMPLETION OF THE PHACOEMULSIFICATION PROCEDURE. NO ADDITIONAL COMPLICATION(S) AND/OR MEDICAL INTERVENTION WERE REPORTED. THE PATIENT WAS SUCCESSFULLY IMPLANTED WITH AN INTRAOCULAR LENS AND THE SURGERY ENDED WITH NO OTHER COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
15451 CATALYS PRECISION LASER SYSTEM CATALYS OOE OPTIMEDICA CORPORATION CATALYS-U

Patients

Seq Age Sex Outcome Treatment
1 53 YR Other