FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2914966
·
Received December 20, 2012
Report
- Report Number
- 1720753-2012-10692
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- December 7, 2012
- Report Date
- December 20, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CONNECTOR AND THE VOLTAGE ON THE 5V POWER SUPPLY WERE ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CONTROL PANEL ERROR MESSAGE AND WOULD NOT TAKE IMAGES. THERE IS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |