FDA Adverse Event Summary report: N

BRIGHTVIEW XCT IMAGING SYSTEM

MDR report key: 2914853 · Received January 10, 2013

Report

Report Number
1525965-2013-00006
Date Received
January 10, 2013
Report Date
December 12, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Product Code
KPS
PMA / PMN Number
K080927
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

PHILIPS RECEIVED INFORMATION WHERE THE CUSTOMER DURING MORNING QC, DEPLOYED THE FLAT PANEL FROM THE SYSTEM AND EXPERIENCED RESISTANCE WHEN ATTEMPTING TO ENGAGE THE LOCKING HANDLE. THE CUSTOMER REPORTED THE LOCKING HANDLE APPEARED TO BE IN PLACE. THE CUSTOMER THEN ROTATED THE GANTRY TO START ANGLE AND THE FLAT PANEL BECAME DISLODGED AND FELL INTO DETECTOR. THE CUSTOMER STATED THE SYSTEM DID NOT INDICATE THAT THE FLAT PANEL WAS NOT IN THE LOCKED POSITION. THERE WAS NO PATIENT INVOLVEMENT AND THERE WAS NO REPORT OF HARM TO THE OPERATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
14987 BRIGHTVIEW XCT IMAGING SYSTEM KPS PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. 882482

Patients

Seq Age Sex Outcome Treatment
1