FDA Adverse Event
Summary report: N
BRIGHTVIEW XCT IMAGING SYSTEM
MDR report key: 2914853
·
Received January 10, 2013
Report
- Report Number
- 1525965-2013-00006
- Date Received
- January 10, 2013
- Report Date
- December 12, 2012
- Manufacturer
- PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
- Product Code
- KPS
- PMA / PMN Number
- K080927
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).
Description of Event or Problem · 1
PHILIPS RECEIVED INFORMATION WHERE THE CUSTOMER DURING MORNING QC, DEPLOYED THE FLAT PANEL FROM THE SYSTEM AND EXPERIENCED RESISTANCE WHEN ATTEMPTING TO ENGAGE THE LOCKING HANDLE. THE CUSTOMER REPORTED THE LOCKING HANDLE APPEARED TO BE IN PLACE. THE CUSTOMER THEN ROTATED THE GANTRY TO START ANGLE AND THE FLAT PANEL BECAME DISLODGED AND FELL INTO DETECTOR. THE CUSTOMER STATED THE SYSTEM DID NOT INDICATE THAT THE FLAT PANEL WAS NOT IN THE LOCKED POSITION. THERE WAS NO PATIENT INVOLVEMENT AND THERE WAS NO REPORT OF HARM TO THE OPERATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 14987 | BRIGHTVIEW XCT IMAGING SYSTEM | KPS | PHILIPS MEDICAL SYSTEMS(CLEVELAND), INC. | 882482 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |