KEEPSAFE DELUXE
Report
- Report Number
- 2020362-2012-00673
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Report Date
- November 16, 2012
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
RESULTS: EVAL OF THE RETURNED UNIT DID NOT CONFIRM THE REPORTED ISSUE OF NO POWER. THE UNIT POWERED ON WHEN USING THE CUSTOMER'S BATTERIES OR AN EXTERNAL POWER SUPPLY. THERE IS BATTERY LEAKAGE RESIDUE ON THE BATTERY DIAGRAM INSIDE BATTERY COMPARTMENT. NO VISIBLE LEAKAGE OR CORROSION TO THE BATTERY SPRINGS OR FLAT CONTACTS. UNIT WAS TESTED A TOTAL OF THREE TIMES AT FIVE MINUTE INTERVALS AND NO INTERMITTENCY WAS DETECTED. UNIT PASSES ALL FUNCTIONAL TESTS. NOTE: INSTRUCTIONS FOR USE STATE: TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. ALWAYS INSTALL A COMPLETELY NEW SET OF BATTERIES WHEN THE CHIRPING DUE TO LOW BATTERY POWER SOUND BEGINS. DO NOT REPLACE A SINGLE CELL, BUT ALL CELLS IN THE ALARM. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK. USE OF MIXED BATTERIES OR BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. REMOVE ANY ALARM FROM USE AND SEND TO THE APPROPRIATE FACILITY AUTHORITY IF BATTERIES ARE DAMAGED OR CORRODED OR THE BATTERY COMPARTMENT HAS SIGNS OF PREVIOUS BATTERY CORROSION SUCH AS WHITE POWDER RESIDUE. (B)(4).
CUSTOMER REPORTED THE ALARM HAS NO POWER. BATTERIES HAVE BEEN REPLACED WITH A NEW SUPPLY ALL OF THE SAME TYPE. NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM. NO PT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KEEPSAFE DELUXE | KMI | J. T. POSEY CO. | 8374 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | OVER MATTRESS SENSOR PART #8307, LOT #UNK |