FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2914538 · Received January 14, 2013

Report

Report Number
2954323-2013-00021
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 18, 2012
Report Date
December 21, 2012
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 20 MG/DL TO 500 MG/DL. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 500 MG/DL. A BLOOD GLUCOSE READING ABOVE 500 MG/DL WILL READ AS A "HI" IN THE ADC METER.

Additional Manufacturer Narrative · 1

REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (1267251) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF "HI" (GREATER THAN 500 MG/DL), 98 MG/DL, 379 MG/DL AND 109 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20426 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1267251

Patients

Seq Age Sex Outcome Treatment
1 31 YR