FDA Adverse Event Malfunction Summary report: N

INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2914456 · Received January 14, 2013

Report

Report Number
9611451-2013-00040
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 21, 2012
Report Date
December 21, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT235 INFANT BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED INFANT BREATHING CIRCUIT. THE PRESSURE TEST RESULT WAS WITHIN THE REQUIRED SPECIFICATION. A LOT CHECK WAS NOT PERFORMED AS NO LOT INFORMATION HAD BEEN PROVIDED. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED BREATHING CIRCUIT AND THE REPORTED FAULT WAS NOT REPLICATED DURING TESTING. ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS PRIOR TO LEAVING THE PRODUCTION LINE, AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARM."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20745 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT235 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1