ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-00893
- Event Type
- Injury
- Date Received
- January 14, 2013
- Date of Event
- December 24, 2012
- Report Date
- January 16, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE LEAD WAS RETURNED FOR TESTING AND IS CURRENTLY BEING EVALUATED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY. THIS PRODUCT ISSUE WILL BE UPDATED WHEN EVALUATION IS COMPLETE.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD HAD DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS REMOVED FROM SERVICE AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19709 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0295 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization | E102| 0157| T135| 0295| 0296| E142| 4076| 4136 |