FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX
MDR report key: 2914196
·
Received December 7, 2012
Report
- Report Number
- 1218950-2012-04040
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Report Date
- November 15, 2012
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED THAT THE CHARGE BUTTON FAILED INTERMITTENTLY IN OPERATIONAL CHECK. THERE WAS NO PT INVOLVEMENT. THE DEVICE WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS VERIFIED. THE ISSUE WAS LOCALIZED TO THE PROCESSOR PCA. THE PROCESSOR PCA WAS REPLACED TO RESOLVE THE ISSUE. THE DEVICE PASSED ALL TESTING AND WAS SHIPPED BACK TO THE CUSTOMER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE CHARGE BUTTON FAILED INTERMITTENTLY IN OPERATIONAL CHECK. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS MEDICAL SYSTEMS | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |