FDA Adverse Event
Malfunction
Summary report: N
LINOX SD 65/16
MDR report key: 2914104
·
Received December 7, 2012
Report
- Report Number
- 1028232-2012-03163
- Event Type
- Malfunction
- Date Received
- December 7, 2012
- Report Date
- November 28, 2012
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- LWS
- PMA / PMN Number
- P980023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MFR OF THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BENN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
Description of Event or Problem · 1
OUS MDR - AFTER AN IMPLANT DURATION OF ABOUT 45 MONTHS HIGH PACING IMPEDANCES (OVER 2000 OHMS) AND LOSS OF CAPTURED WERE NOTED. NO ADVERSE PT SIDE EFFECTS WERE REPORTED. THE LEAD WAS NOT EXPLANTED. THE EVENT DATE WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINOX SD 65/16 | ICD LEAD | LWS | BIOTRONIK SE & CO. KG | 350053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |