FDA Adverse Event Malfunction Summary report: N

LINOX SD 65/16

MDR report key: 2914104 · Received December 7, 2012

Report

Report Number
1028232-2012-03163
Event Type
Malfunction
Date Received
December 7, 2012
Report Date
November 28, 2012
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
LWS
PMA / PMN Number
P980023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MFR OF THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BENN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.

Description of Event or Problem · 1

OUS MDR - AFTER AN IMPLANT DURATION OF ABOUT 45 MONTHS HIGH PACING IMPEDANCES (OVER 2000 OHMS) AND LOSS OF CAPTURED WERE NOTED. NO ADVERSE PT SIDE EFFECTS WERE REPORTED. THE LEAD WAS NOT EXPLANTED. THE EVENT DATE WAS NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINOX SD 65/16 ICD LEAD LWS BIOTRONIK SE & CO. KG 350053

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization