FDA Adverse Event Injury Summary report: N

REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 10

MDR report key: 2914006 · Received January 14, 2013

Report

Report Number
0002249697-2013-00044
Event Type
Injury
Date Received
January 14, 2013
Date of Event
December 31, 2012
Report Date
December 31, 2012
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K092561
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT IS (B)(6). AN EVENT REGARDING CORROSION INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. REVIEW OF DEVICE HISTORY RECORDS FOUND THE DEVICES IN THE REPORTED LOT WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE SUPER AND CHS COMPLAINT DATABASES SHOW THERE HAVE BEEN OTHER EVENTS FOR THE REPORTED LOT. SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED CORROSION IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. (B)(4): NOT RETURNED TO MANUFACTURER.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON REVISED PATIENT'S RIGHT REJUVENATE HIP DUE TO CORROSION AT THE NECK STEM JUNCTION. THE CUP REMAINED IN AS SURGEON SAID IT WAS WELL FIXED. SURGEON REPLACED WITH A RESTORATION MODULAR STEM.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON REVISED PATIENT'S RIGHT REJUVENATE HIP DUE TO CORROSION AT THE NECK STEM JUNCTION. THE CUP REMAINED IN AS SURGEON SAID IT WAS WELL FIXED. SURGEON REPLACED WITH A RESTORATION MODULAR STEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21435 REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 10 IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH MHTP6X

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention