FDA Adverse Event Malfunction Summary report: N

STRESSWRITER EXERCISE TESTING SYSTEM

MDR report key: 291399 · Received August 15, 2000

Report

Report Number
1218950-2000-00051
Event Type
Malfunction
Date Received
August 15, 2000
Date of Event
June 6, 2000
Report Date
June 7, 2000
Manufacturer
AGILENT TECHNOLOGIES, INC.
Product Code
DQK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS STARTED IN PRE-EXERCISE MODE (DEFAULT CONFIGURED AT 0.8 KM/H AND 0%). WHEN PT WAS WALKING ON THE TREADMILL, SUDDENLY THE EXERCISE DEVICE STARTED TO RUN A MAXIMUM SPEED (16, 1 KM/H) OUT OF CONTROL (BRUCE PROTOCOL WAS RUNNING). "PC" KEYBOARD DIDN'T RESPOND TO A STOP COMMAND; INSTEAD THE TREADMILL EMERGENCY STOP SWITCH, WHICH WOULDN'T BE REACHED BY THE PT, HAD TO BE PRESSED BY THE DR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRESSWRITER EXERCISE TESTING SYSTEM STRESS TEST SYSTEM DQK AGILENT TECHNOLOGIES, INC. M2488A NA

Patients

Seq Age Sex Outcome Treatment
1 *