FDA Adverse Event
Malfunction
Summary report: N
DIGITAL RADIOGRAPHY CXDI-70C WIRELESS
MDR report key: 2913815
·
Received September 25, 2012
Report
- Report Number
- 1000181430-2012-00042
- Event Type
- Malfunction
- Date Received
- September 25, 2012
- Report Date
- July 25, 2012
- Manufacturer
- CANON, INC.
- Product Code
- MQB
- PMA / PMN Number
- K102012
- Removal / Correction Number
- Z-2265-2012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION OF THE RETURNED UNIT CONFIRMED THAT THE PANEL HAD LOOSE SCREWS INSIDE. THERE WAS A BURNT (SHORTED) AREA ON THE DC/DC PC BOARD. THE DC/DC PC BOARD WAS REPLACED. THE UNIT WAS SERVICED FOR BOTH LOOSE SCREWS INSIDE THE UNIT AND THE PANEL COVER SCREWS. THE SERVICE REPRESENTATIVE PERFORMED CALIBRATION AND SELF-TESTS, AND THE UNIT MET SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE UNIT WOULD NOT POWER UP AND THEY CAN HEAR RATTLING INSIDE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIGITAL RADIOGRAPHY CXDI-70C WIRELESS | MQB | CANON, INC. | CXDI-70C WIRELESS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |