FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2913739 · Received January 14, 2013

Report

Report Number
2124215-2013-00177
Event Type
Injury
Date Received
January 14, 2013
Date of Event
December 30, 2012
Report Date
January 24, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY IT WAS CONFIRMED THAT THE LEAD APPEARED KINKED AND SHOWED SIGNS OF HAVING BEEN TWISTED WHICH IS CONSISTENT WITH TWIDDLER'S SYNDROME. TWIDDLER'S SYNDROME CAN LEAD TO DISLODGEMENT AND COMPLICATIONS RELATED TO DISLODGEMENT.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION REVEALS THAT THIS PATIENT EXHIBITED TWIDDLER'S SYNDROME AND UNDERWENT A REVISION PROCEDURE DUE TO DISLODGEMENT. THE LEAD WAS EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD DISLODGED. THE PATIENT WAS ADMITTED INTO THE HOSPITAL AND WAS FEELING SOME SPASMS IN THEIR STOMACH. THERE WAS A NOTED DROP IN P-WAVES FROM 4.5 TO 0.3, AN INCREASE IN IMPEDANCES TO 620 OHMS AND NON-CAPTURE AT 5 VOLTS AT 2 MILLISECONDS. IT WAS REPORTED THAT THE PHYSICIAN HAD THE PATIENT'S GENERATOR PROGRAMMED TO PACE IN THE VENTRICLE ONLY AND THE PATIENT WAS DISCHARGED AND WAS TO BE RE-EVALUATED IN THE CLINIC AT A LATER DATE. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19694 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4470

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| L 0296| E142| 4470