CONTAK RENEWAL
Report
- Report Number
- 2124215-2012-16997
- Event Type
- Malfunction
- Date Received
- January 14, 2013
- Date of Event
- December 19, 2012
- Report Date
- January 4, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
ANALYSIS WAS PERFORMED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THIS DEVICE HAD A COMPROMISED LOW-VOLTAGE CAPACITOR THAT DID NOT MEET DESIGN SPECIFICATION. DESPITE THIS COMPROMISED CAPACITOR, CURRENT LEAKAGE WAS NOT SUFFICIENT TO ADVERSELY IMPACT DEVICE LONGEVITY. NORMAL PACING, SENSING, AND DEFIBRILLATION THERAPY FUNCTIONS WERE VERIFIED DURING TESTING.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS RETURNED WITH NO REPORTED PRODUCT PERFORMANCE ISSUES AND NO REPORTED ADVERSE PATIENT EFFECTS. INITIAL ANALYSIS COMPLETED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY DETERMINED THIS PRODUCT DID NOT MEET LONGEVITY EXPECTATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 20594 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H195 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | H195 |