FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2913682 · Received January 14, 2013

Report

Report Number
2124215-2012-16997
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 19, 2012
Report Date
January 4, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

ANALYSIS WAS PERFORMED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THIS DEVICE HAD A COMPROMISED LOW-VOLTAGE CAPACITOR THAT DID NOT MEET DESIGN SPECIFICATION. DESPITE THIS COMPROMISED CAPACITOR, CURRENT LEAKAGE WAS NOT SUFFICIENT TO ADVERSELY IMPACT DEVICE LONGEVITY. NORMAL PACING, SENSING, AND DEFIBRILLATION THERAPY FUNCTIONS WERE VERIFIED DURING TESTING.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS RETURNED WITH NO REPORTED PRODUCT PERFORMANCE ISSUES AND NO REPORTED ADVERSE PATIENT EFFECTS. INITIAL ANALYSIS COMPLETED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY DETERMINED THIS PRODUCT DID NOT MEET LONGEVITY EXPECTATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20594 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H195

Patients

Seq Age Sex Outcome Treatment
1 H195