FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2913667 · Received January 14, 2013

Report

Report Number
2124215-2013-00006
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 27, 2012
Report Date
April 10, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES [, AND THE ERI TO EOL TIME PERIOD WAS SHORTENED,] DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

INFORMATION WAS SUBSEQUENTLY RECEIVED THAT THIS DEVICE WAS EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED AS A RESULT OF THE PROCEDURE.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT PRESENTED TO THE CLINIC AS THE DEVICE WAS BEEPING. UPON INTERROGATION, THE DEVICE HAD DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI). IT WAS NOTED THAT THE BATTERY VOLTAGE WAS 2.66 VOLTS, HOWEVER, THE CHARGE TIME WAS 20 SECONDS. A REVISION PROCEDURE WAS SCHEDULED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20221 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 50 YR T177| MISMATCH