FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT

MDR report key: 2913652 · Received January 14, 2013

Report

Report Number
3005075853-2013-00210
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 18, 2012
Report Date
December 19, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). ATTEMPTS HAVE BEEN MADE TO HAVE DEVICE RETURNED. DEVICE LOCATION IS UNKNOWN. THE DEVICE WAS NOT RECEIVED FOR ANALYSIS; THEREFORE, THE COMPLAINT COULD NOT BE CONFIRMED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE BLADE CAME OFF THE RAIL (WENT FROM OUTSIDE TO INSIDE), THEN FELL APART. USED SAME LIKE PRODUCT TO COMPLETE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21595 PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFTEY LOCK-OUT STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4CH7C

Patients

Seq Age Sex Outcome Treatment
1