TELIGEN
Report
- Report Number
- 2124215-2013-00076
- Event Type
- Malfunction
- Date Received
- January 14, 2013
- Date of Event
- December 19, 2012
- Report Date
- January 31, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Removal / Correction Number
- Z-0026-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF THE DEVICE MEMORY CONFIRMED THE LOW VOLTAGE FAULT. THE FAULT WAS DECLARED ON (B)(4), 2012 DUE TO THREE DAILY VOLTAGES THAT WERE BELOW THRESHOLD. THERE WERE NO OTHER SUSPECIOUS FAULTS OR RESETS STORED WITHIN THE DEVICE MEMORY. BASED ON THE BATTERY VOLTAGE MEASUREMENT, THERAPY DELIVERY WOULD NOT HAVE BEEN COMPROMISED. HOWEVER, SINCE THE DEVICE HARDWARE IS NOT DETECTING THE LOSS OF BATTERY ENERGY, CAUSING THE INDICATORS TO BE INACCURATE, A REPLACEMENT WAS RECOMMENDED AS SOON AS POSSIBLE.THE DEVICE REMAINS IN SERVICE AT THIS TIME. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
(B)(4). BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO THIS ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.
APPROXIMATELY THREE WEEKS LATER, THE DEVICE WAS EXPLANTED AND REPLACED DUE TO THE REPORTED OBSERVATIONS. THE PRODUCT WAS RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. REVIEW OF THE DEVICE MEMORY CONFIRMED THREE LOW VOLTAGE BATTERY FAULTS (FAULT CODE 1003) HAD BEEN RECORDED. EXTERNAL VISUAL INSPECTION OF THE DEVICE NOTED BODY FLUID CONTAMINATION IN THE DEFIBRILLATION (POSITIVE) LEAD BARREL AND TERMINAL BLOCK. THE BATTERY WAS THEN REMOVED, SO THAT AN EXTERNAL POWER SUPPLY COULD BE CONNECTED TO THE DEVICE FOR FURTHER ANALYSIS. A HIGHER THAN NORMAL CURRENT USAGE WAS OBSERVED. ELECTRICAL TESTING ISOLATED THE HIGH CURRENT CONDITION TO TWO COMPROMISED LOW VOLTAGE CAPACITORS THAT ARE CONNECTED TO THE DEVICE'S BATTERY. THIS TYPE OF COMPONENT MALFUNCTION CAN RESULT IN A HIGH CURRENT DRAIN, WHICH OVER TIME CAN RESULT IN PREMATURE BATTERY DEPLETION, AS WAS OBSERVED IN THIS CASE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EMITTING BEEP TONES SO THE PATIENT WAS BROUGHT IN TO THE CLINIC FOR A DEVICE CHECK. UPON DEVICE INTERROGATION, A FAULT CODE WAS DISPLAYED INDICATING THE BATTERY VOLTAGE WAS TOO LOW FOR THE PROJECTED REMAINING CAPACITY; HOWEVER THE DEVICE LONGEVITY SHOWED 9.5 YEARS REMAINING. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) WAS CONTACTED AND DISCUSSED THAT THERE WAS A SOURCE OF HIGH CURRENT DRAIN OCCURRING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 21421 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWP | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | E102| 0175 |