FDA Adverse Event Other Summary report: N

DCA VANTAGE

MDR report key: 2913620 · Received January 7, 2013

Report

Report Number
1217157-2012-00075
Event Type
Other
Date Received
January 7, 2013
Date of Event
November 27, 2012
Report Date
November 27, 2012
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS
Product Code
CGX
PMA / PMN Number
K071466
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER WILL IDENTIFY REFERENCE METHOD USED BY (B)(6). MFR INVESTIGATING.

Description of Event or Problem · 1

CUSTOMER REPORTS DISCORDANT HBA1C RESULTS ON TWO KNOWN DIABETIC PTS. PT 1: PT PRESENTS AS (B)(6) KNOWN DIABETIC. DIABETES TREATED WITH STARLIX (NATEGLINIDE). PT HAD A RESULT OF 6.0% ON THE DCA VANTAGE VIA FINGERSTICK, STATES HE HAD A1C PERFORMED BY (B)(6) ON (B)(6) 2012 VIA VENIPUNCTURE, RESULT OF 7.7%. RESULT OF 6.0% AND 7.7% WERE GIVEN TO THE PHYSICIAN, UNK IF ACTION WAS TAKEN ON EITHER RESULT. OTHER MEDICATIONS INCLUDE LIPITOR, DIOVAN, CLONIDINE AND ATENOLOL. PREVIOUS A1C RESULTS: (B)(6) 2012: 6.3% (REFERENCE LAB). (B)(6) 2012: 5.6% (REFERENCE LAB).

Additional Manufacturer Narrative · 2

CUSTOMER WILL IDENTIFY REFERENCE METHOD USED BY (B)(6). MFR INVESTIGATING.

Description of Event or Problem · 2

CUSTOMER REPORTS DISCORDANT HBA1C RESULTS ON TWO KNOWN DIABETIC PTS. PT 2: PT 2 PRESENTS AS A KNOWN DIABETIC FEMALE, AGE UNK, WEIGHT UNK. NOT PREGNANT. RESULT OF 6.0% ON DCA VANTAGE, REFERENCE RESULT OF 8.1%. REFERENCE METHODOLOGY UNK. MEDICATION UNK. NO HISTORICAL RESULTS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8409 DCA VANTAGE DCA VANTAGE CGX SIEMENS HEALTHCARE DIAGNOSTICS VANTAGE

Patients

Seq Age Sex Outcome Treatment
1 62 YR
2 UNK