FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2913505 · Received January 9, 2013

Report

Report Number
8020893-2013-00066
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
December 13, 2012
Report Date
December 13, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD AN ERRATIC SCREEN. THERE WAS NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
13152 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCO 840

Patients

Seq Age Sex Outcome Treatment
1