FOLFUSOR

MDR report key: 2913456 · Received January 14, 2013

Report

Report Number
1416980-2013-00873
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 1, 2012
Report Date
December 19, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS PRODUCT IS NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510(K) NUMBER.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING APPROXIMATELY 60 ML OF FLUID IN THE RESERVOIR. THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGNS OF PHYSICAL ABNORMALITY. A LEAK TEST SHOWED NO SIGNS OF LEAKAGE ON ANY PART OF THE DEVICE. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT A FOLFUSOR LEAKED AFTER FILLING. THE DEVICE WAS FILLED WITH A SOLUTION OF 5-FU. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21062 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12F066

Patients

Seq Age Sex Outcome Treatment
1