ACUITY
Report
- Report Number
- 2124215-2012-17243
- Event Type
- Injury
- Date Received
- January 14, 2013
- Date of Event
- December 10, 2012
- Report Date
- January 14, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE (LOC) APPROXIMATELY TWO MONTHS POST IMPLANT. THE LEAD HAD DISLODGED. A REVISION PROCEDURE WAS PERFORMED. THE LEAD WAS EXPLANTED. ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19786 | ACUITY | IMPLANTABLE HF LEAD | OJX | CPI - DEL CARIBE | 4592 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Hospitalization| L| R | N160| 4592| 0295| 4470 |