FDA Adverse Event
Injury
Summary report: N
FINELINE II
MDR report key: 2913236
·
Received January 14, 2013
Report
- Report Number
- 2124215-2012-17480
- Event Type
- Injury
- Date Received
- January 14, 2013
- Date of Event
- December 10, 2012
- Report Date
- December 10, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- PMA / PMN Number
- P960004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SUCCESSFULLY REPOSITIONED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ATRIAL LEAD WAS EXHIBITING VARYING THRESHOLD AND IMPEDANCE MEASUREMENTS ONE WEEK POST IMPLANT. AN X-RAY REVEALED THE LEAD HAD DISLODGED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19785 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Hospitalization | 0276| E142| 4470 |